Proper Cleaning & Storage Tips For Reusable Surgical Gel Position Pads
Part 1: Step-by-Step Standard Cleaning & Disinfection Workflow
- Initial Dry Wipe Removal
Use disposable lint-free cloths to wipe off visible contaminants: blood, antiseptic solution, irrigation fluid, bodily secretions. Remove debris before applying any disinfectant.
- Select Compatible Disinfectants
✅ Recommended (Low / Intermediate-level disinfection):
- 70% Isopropyl alcohol wipes
- EPA-registered quaternary ammonium compound (quat) wipes
- Diluted hospital-grade tuberculocidal disinfectants
- High-level disinfectants: glutaraldehyde, OPA, concentrated peracetic acid
- Bleach in high concentrations
- Abrasive scouring powders, steel wool, hard bristle brushes
- Steam cleaning, autoclaving and heat sterilisation
Harsh chemicals damage PU/TPU outer film, causing stickiness, microcracks and seam failure.
- Wipe Entire Surface & Seams Thoroughly
Gently wipe top surface, sides, bottom anti-slip layer and all welded seams. Seams are the most vulnerable area for fluid infiltration. Maintain the required contact time specified by the disinfectant manufacturer.
- Complete Air Drying Before Next Use
Do not reuse damp gel pads. Moisture trapped against the cover accelerates material degradation and encourages microbial growth. Place pads flat in a well-ventilated area until fully dry.
- Visual Inspection After Every Cleaning
Check for these defects; isolate and remove damaged pads immediately:
- Surface stickiness, cloudiness or fine cracks
- Bubbling, peeling or separation of outer film
- Discolouration around welded seams
- Any signs of gel leakage
Part 2: Common Harmful Cleaning Mistakes To Avoid
- Scrubbing aggressively with rough brushes
Creates micro-abrasions on the waterproof film, enabling fluid and bacteria to penetrate.
- Mixing different types of disinfectants
Chemical reactions can permanently damage the cover material and leave toxic residues.
- Leaving disinfectant residue to air-dry without wiping
Residual chemicals gradually break down TPU/PU material over repeated exposure.
- Stacking wet gel pads together
Traps moisture, speeds up mould formation and causes surface adhesion.
- Attempting immersion cleaning
Even sealed pads are not designed for soaking; water can slowly penetrate minor invisible seam defects.
Part 3: Best Practices For Storage Of Reusable Gel Position Pads
General Storage Rules
- Store pads in a clean, dry, well-ventilated OR supply room, away from direct sunlight, UV light and heating units. Continuous heat and UV rays accelerate polymer aging.
- Do not fold or crease contoured gel pads long-term
Prolonged sharp folds create permanent creases and internal gel separation. Flat storage is ideal.
- Avoid heavy objects stacked on top of gel wedges and mats
Sustained pressure causes permanent compression set, destroying pressure redistribution performance.
- Separate different gel positioners; prevent sharp surgical instruments from contacting pad surfaces. Tiny punctures lead to gel leakage.
Two Practical Storage Methods
- Flat shelf storage (Preferred for large torso mats, headrests)
Lay horizontally on clean open shelves, single layer. Simple, lowest risk of deformation.
- Upright rack storage (For wedges, axillary rolls, heel supports)
Use dedicated wire racks to stand contoured positioners on edge. Prevent chaotic piling.
Long-Term Unused Storage
- Fully clean, disinfect and air-dry completely first
- Loosely place inside breathable non-woven storage bags (avoid airtight plastic bags that trap humidity)
- Inspect every 30 days for discolouration or mould
Part 4: Department-Specific Extra Maintenance Tips
Obstetric Suites
Plastic & Aesthetic Surgery ORs
Radiology & CT Departments
Isolation Rooms
Part 5: Establish Replacement Timelines
- Hospital-grade well-maintained gel positioners: 3–5 years
- Pads exposed to frequent heavy contamination or intermediate disinfection: expect shorter lifespan
Do not continue using damaged pads as temporary stopgaps; leakage creates sterility and infection hazards.
Frequently Asked Questions
Q1: Can I use spray disinfectants on gel position pads?
Q2: Is mould on gel pads repairable?
Q3: Can gel pads be transported in sealed plastic bins between operating rooms?
Closing Paragraph
Routine proper cleaning, validated disinfectant selection and compliant storage directly extend the lifespan of reusable surgical gel position pads. Building clear standard operating procedures for OR staff reduces premature equipment failure, controls procurement expenditure and maintains consistent infection control standards.
Train clinical teams on prohibited chemicals and storage rules to protect your facility’s investment in positioning equipment. Reach out to our team to receive printable maintenance guidance documents and bulk supply solutions for medical gel positioners.